2. Introduction
The Fingertip pulse oximeter is to spot-check oxygen saturation in
blood (SpO
2
) and pulse rate. The pulse oximeter is used on adults
and pediatric at hospital, clinics, and/or home.
The device contains a dual light source (Red LED and Infrared red
LED) and a photodetector. Bone, tissue, pigmentation, and venous
vessels normally absorb a constant amount of light over time. The
arteriolar bed normally pulsates and absorbs variable amounts of
light during the pulsations. The ratio of light absorbed is translated in
an oxygen saturation measurement (SpO
2
). Because a
measurement of SpO
2
is dependent on light from the device,
excessive ambient light can interfere with this measurement.
The wavelength of red LED is 660nm and Infrared LED is
905/880nm with maximum optical output power of 4mW.
3. Features
Measure and display reliable SpO
2
value and heart rate.
Single turn-on key for easy operation.
Bright Organic LED display.
Light, compact, and portable.
Battery power is for a continuing use of 16 hours.
Two AAA Alkaline batteries for power supply.
Device will be off automatically after 8 seconds in idling.
Visual alarm (SB200 except) and audio alarm (SB220 only).